Duodenoscope Sheath Protects High-Risk End of Instrument
Bret T. Petersen, MD, FASGE, reviewing Pasricha PJ, et al. Gastrointest Endosc 2020 Mar 6.
The risk of transmission of microorganisms by contaminated duodenoscopes remains a pertinent concern for ERCP practices. This report describes microbial challenge studies in a single-use device recently cleared by the Food and Drug Administration (FDA). The device, analogous to a condom, appears to effectively prevent “outside-in” microbial contamination of the high-risk elevator region and the leading end of the endoscope and device channel. The same device also provides “inside-out” protection by sealing in existing contaminants, preventing potential transmission to the patient. Bench studies confirmed that mechanical and optical performance of the endoscope was not compromised by the device.
Note to readers: At the time we reviewed this paper, its publisher noted that it was not in final form and that subsequent changes might be made.
CITATION(S)
Pasricha PJ, Miller S, Carter F, Humphries R. A novel and effective disposable device that provides 2-way protection to the duodenoscope from microbial contamination. Gastrointest Endosc 2020 Mar 6. (Epub ahead of print) (https://doi.org/10.1016/j.gie.2020.03.001)